Platelet Count Pitfalls Between Routine and Advanced Techniques
Main Article Content
Keywords
Platelet enumeration, automated hematology analyzers, flow cytometry validation.
Abstract
Background: Precise quantification of platelet counts is a fundamental aspect of clinical hematology, pivotal in diagnosing and managing conditions such as thrombocytopenia and thrombocytosis. However, inconsistencies in platelet measurements across various automated hematology analyzers present significant challenges, potentially compromising the reliability of results. This variability is especially critical when platelet counts inform urgent clinical interventions, including the administration of platelet transfusions. Aim: The present study evaluated the accuracy of Sysmex XN-350 and Mindray in comparison to flow cytometry, the gold standard for platelet counting. Methods: 120 blood samples were categorized into thrombocytopenic, thrombocytosis, and normal groups. Platelet counts were measured using Sysmex XN-350, Mindray BC-720, and flow cytometry. Statistical analyses, including correlation coefficients, Bland-Altman plots, and repeated measures ANOVA, were employed to assess agreement and differences among the methods. Results: The study revealed significant discrepancies in platelet counts among the devices. Sysmex XN-350 consistently overestimated platelet counts compared to flow cytometry, particularly at higher counts, with a mean difference of 176.76 ± 358.51. Mindray BC-720 demonstrated greater agreement with flow cytometry, with a mean difference of 60.062 ± 119.67 and a stronger correlation (r = 0.972). Bland-Altman analysis showed that Sysmex exhibited substantial overestimation at higher platelet counts, while Mindray maintained consistency within clinically relevant ranges. Conclusions: Mindray BC-720 outperformed Sysmex XN-350 in terms of agreement with flow cytometry, especially in the thrombocytopenic and normal ranges. These findings highlight the importance of validating automated hematology analyzers against advanced techniques like flow cytometry to ensure precision in platelet enumeration.
References
2. Thachil J, Bates I. Approach to the diagnosis and classification of blood cell disorders. Dacie and Lewis Practical Haematology. 2017:497.
3. Hvas A-M, editor. Crucial stepping stones in platelet history. 2023: Thieme Medical Publishers, Inc.
4. Baccini V, Geneviève F, Jacqmin H, Chatelain B, Girard S, Wuilleme S, et al. Platelet Counting: Ugly Traps and Good Advice. Proposals from the French-Speaking Cellular Hematology Group (GFHC). J Clin Med. 2020;9(3).
5. Liao K, Zhang X, Liu J, Teng F, He Y, Cheng J, et al. The role of platelets in the regulation of tumor growth and metastasis: the mechanisms and targeted therapy. MedComm. 2023;4(5):e350.
6. Oh SE, Seo JE, An JY, Lee JH, Sohn TS, Bae JM, et al. Prognostic impact of increased perioperative platelet count in gastric cancer patients. Journal of Surgical Research. 2019;242:296-303.
7. Diks AM, Bonroy C, Teodosio C, Groenland RJ, De Mooij B, De Maertelaere E, et al. Impact of blood storage and sample handling on the quality of high-dimensional flow cytometric data in multicenter clinical research. Journal of Immunological Methods. 2019;475:112616.
8. Baccini V, Geneviève F, Jacqmin H, Chatelain B, Girard S, Wuilleme S, et al. Platelet counting: ugly traps and good advice. Proposals from the French-Speaking Cellular Hematology Group (GFHC). Journal of Clinical Medicine. 2020;9(3):808.
9. Green R, Wachsmann-Hogiu S. Development, history, and future of automated cell counters. Clinics in Laboratory Medicine. 2015;35(1):1-10.
10. Muturi CK. Comparative assessment of automated and manual white blood cell counts at Kenyatta National Hospital Haematology Laboratory. 2008.
11. Hemoglobinometry A, Red C, Histogram E, Width RCD. Principles and Practice of Clinical Hematology. Linne & Ringsrud's Clinical Laboratory Science-E-Book: The Basics and Routine Techniques. 2015;2:291.
12. Boulassel MR, Al-Farsi R, Al-Hashmi S, Al-Riyami H, Khan H, Al-Kindi S. Accuracy of Platelet Counting by Optical and Impedance Methods in Patients with Thrombocytopenia and Microcytosis. Sultan Qaboos University Medical Journal. 2015;15(4):e463-8.
13. Sun Y, Hu Z, Huang Z, Chen H, Qin S, Jianing Z, et al. Compare the accuracy and precision of Coulter LH780, Mindray BC-6000 Plus, and Sysmex XN-9000 with the international reference flow cytometric method in platelet counting. PLoS One. 2019;14(5):e0217298.
14. Hummel K, Sachse M, Hoffmann JJML, van Dun LPJM. Comparative evaluation of platelet counts in two hematology analyzers and potential effects on prophylactic platelet transfusion decisions. Transfusion. 2018;58(10):2301-8.
15. Al–Saeed AHM. Evaluation of oxidative status, potassium, magnesium, and lipid profile in serum of patients with β-thalassemia major, Thi-Qar, Iraq. Maaen Journal for Medical Sciences. 2023;2(3):3.
16. Cook AM, Moritz A, Freeman KP, Bauer N. Objective evaluation of analyzer performance based on a retrospective meta‐analysis of instrument validation studies: point‐of‐care hematology analyzers. Veterinary Clinical Pathology. 2017 Jun;46(2):248-61.
17. Kakkar N. Spurious rise in the automated platelet count because of bacteria. J Clin Pathol. 2004;57(10):1096-7.
18. Shenoy SS. A Study of Automated Complete Blood Count Parameters in Various Hematological Disorders. 2016.
19. Pan L-L, Chen C-M, Huang W-T, Sun C-K. Enhanced accuracy of optical platelet counts in microcytic anemia. Laboratory Medicine. 2014;45(1):32-6.
20. Gulati G, Uppal G, Gong J. Unreliable automated complete blood count results: causes, recognition, and resolution. Annals of Laboratory Medicine. 2022;42(5):515-30.
21. Balakrishnan A, Shetty A, Vijaya C. Estimation of platelet counts: auto analyzer versus counts from peripheral blood smear based on traditional and platelet: red blood cell ratio method. Red. 2018;2(3):4.
22. Wang J, Zhao S, Su Z, Liu X. Analytical comparison between two hematological analyzer systems: Mindray BC-5180 vs. Sysmex XN-1000. J Clin Lab Anal. 2019;33(8):e22955.
23. Gioia M, Da Rin G, Manenti B, Birindelli S, Ciardelli ML, Gentile R, et al. Multicenter evaluation of analytical performances of platelet counts and platelet parameters: carryover, precision, and stability. International Journal of Laboratory Hematology. 2020;42(5):552-64.
24. Kim H, Hur M, Lee G-H, Kim S-W, Moon H-W, Yun Y-M. Performance of platelet counting in thrombocytopenic samples: comparison between Mindray BC-6800Plus and Sysmex XN-9000. Diagnostics. 2021;12(1):68.